AwardedUnited KingdomGoods£56,127
CEFAS25-135 Benchtop carbon analyser.
- Published
- Published 2 February 2026
- Deadline
- Closed 12 January 2026
Overview
1. Technical Requirement
1.1 The instrument must include an autosampler for automated sample handling.
1.2 Suppliers will provide evidence of the instrument's performance range, ideally using marine sediment samples to demonstrate suitability.
1.3 The analysis process will be efficient and cost-effective; please include supporting evidence of throughput and operational costs.
1.4 The instrument will be compatible with other equipment, such as an Isotopic Ratio Mass Spectrometer (IRMS), where possible.
1.5 The instrument must be capable of performing Total Inorganic Carbon (TIC), Total Organic Carbon (TOC), and Residual Oxidisable Carbon (ROC) analyses as a minimum.
1.6 The instrument's maximum dimensions must not exceed 800 mm (W) x 600 mm (D) x 1200 mm (H).
1.7 The instrument's maximum weight must not exceed 85 kg.
1.8 Suppliers must state all gases required for operation, such as nitrogen, oxygen, or synthetic air. There should be no requirement for helium gas.
1.9 The instrument should be supplied with reusable crucibles.
2. Environmental Considerations
2.1 Unit will have a low power consumption.
2.2 Unit will be designed for reuse/remanufacture and of regenerative nature to eliminate waste where possible.
2.3 Consideration of how packaging will be managed. Ideally at the time of installation the supplier will remove packaging and dispose of by ethical and environmentally friendly methods. Where possible, packaging is to be re-used or recycled.
2.4 Consideration of life cycle assessment for the equipment.
3. Hardware, Software, PC, Licenses, operating systems
3.1 The instrument software must be fully compatible with Windows 11.
4. Compatibility with existing hardware/software/equipment
4.1 The instrument will be compatible with the existing Windows 11 PC and Sartorius balance software.
5. Installation
5.1 The supplier must install the instrument and provide all necessary training, which should be included in the total quoted cost.
6. Ease of maintenance
6.1 The bid submission must describe all user-maintainable aspects of the unit, including any limitations where no user-serviceable components are available.
7. Parts, Consumables and Updates
7.1 Crucibles must be reusable, and replacements should be easy to procure.
7.2 Spare parts should be readily available and straightforward to obtain.
7.3 All consumables must be easily accessible.
7.4 The supplier should provide software updates as required.
8. Warranties, Guarantees, Servicing
8.1 Warranties period will start from the point of acceptance of delivery and once the goods have been tested to ensure they are in working order. Cefas expect minimum warranty period of 12 months to be included in the purchase.
Key details
- Country
- United Kingdom
- Status
- Awarded
- Category
- Goods
- Estimated value
- £56,127
- Procedure
- Selective
- SME suitable
- Yes
- Published
- 2 February 2026
- Deadline
- Closed
Gas-analysis apparatus
Award outcome
- Awarded value
- £67,353
- Award date
- 29 January 2026
- Contract period
- 30 January 2026 → 31 March 2027
- Elementar UK Ltd
Contract ending soon? Skim tracks incumbents and flags recompetes early. Book a demo to set up alerts.
Notice history
Every published notice in this contracting process, newest first.
Contract Award Notice
2 February 2026
Get the full picture
AI analysis, bid scoring, and expert strategy for this tender
Skim analyses every tender against your company profile and tells you whether to bid, how to win, and who you are up against.
See which competitors are winning the contracts you're chasing.
15-minute call. We'll pull the win data on your real pipeline and show you what Skim does with it.